Chest

POS. DIST. GRID OR BUCKY MA. MA. for 600 Speed
LAT. 72 YES 200 300
         

System Speeds

200 400 600
TIME MAS. TIME MAS. TIME MAS.
1/10 20 1/15 13.4 1/30 10
 

Thickness of Part in Centimeters

2 4 5 6 10 12 14 15 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46  
- - - - - - - 62 66 70 74 78 82 86 90 94 98 102 106 110 114 - - KV

For single phase equipment 8:1 grid ratio - increase 5 KV for 12:1 grid

 

Contrast Compensating Using Kilovoltage - If you need to decrease the grey scale, please click here.